14.2 : Atomic Spectroscopy: Effects of Temperature

Atomization, converting samples into gas-phase atoms and ions, is essential for atomic spectroscopy. The flame temperature required for atomization affects the efficiency of the atomic spectroscopic methods by increasing the atomization efficiency and the relative population of the excited and ground states.

At thermal equilibrium, the relative populations of excited and ground state atoms can be estimated using the Maxwell–Boltzmann distribution. For example, an increase in temperature from 2500 K to 2600 K can increase the population of excited-state sodium atoms by 45%, while the ground-state population decrease is negligible. Since atomic emission spectroscopy (AES) relies on photon emission from these excited states, it is highly temperature-dependent. In contrast, atomic absorption spectroscopy (AAS) and atomic fluorescence spectroscopy (AFS) primarily depend on the ground-state population and have less significant temperature dependence. However, for easily ionizable elements, an increase in flame temperature causes a loss of atoms by ionization, adversely affecting absorption and fluorescence spectral intensity.

In addition, for atomic spectroscopy overall, higher temperature increases the velocity of the atoms, making the Doppler effect more pronounced. This results in the broadening of atomic spectral lines and decreasing peak height.

Tags

Atomic SpectroscopyAtomizationFlame TemperatureAtomic Emission Spectroscopy AESAtomic Absorption Spectroscopy AASAtomic Fluorescence Spectroscopy AFSMaxwell Boltzmann DistributionExcited state AtomsGround state AtomsTemperature DependenceIonizationSpectral IntensityDoppler EffectSpectral Line Broadening

From Chapter 14:

article

Now Playing

14.2 : Atomic Spectroscopy: Effects of Temperature

Atomic Spectroscopy

267 Views

article

14.1 : Atomic Spectroscopy: Absorption, Emission, and Fluorescence

Atomic Spectroscopy

731 Views

article

14.3 : Atomic Absorption Spectroscopy: Overview

Atomic Spectroscopy

861 Views

article

14.4 : Atomic Absorption Spectroscopy: Instrumentation

Atomic Spectroscopy

500 Views

article

14.5 : Atomic Absorption Spectroscopy: Radiation and Light Sources

Atomic Spectroscopy

309 Views

article

14.6 : Atomic Absorption Spectroscopy: Atomization Methods

Atomic Spectroscopy

341 Views

article

14.7 : Atomic Absorption Spectroscopy: Interference

Atomic Spectroscopy

589 Views

article

14.8 : Atomic Absorption Spectroscopy: Lab

Atomic Spectroscopy

294 Views

article

14.9 : Atomic Emission Spectroscopy: Overview

Atomic Spectroscopy

940 Views

article

14.10 : Atomic Emission Spectroscopy: Instrumentation

Atomic Spectroscopy

310 Views

article

14.11 : Atomic Emission Spectroscopy: Interference

Atomic Spectroscopy

144 Views

article

14.12 : Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle

Atomic Spectroscopy

471 Views

article

14.13 : Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation

Atomic Spectroscopy

171 Views

article

14.14 : Atomic Emission Spectroscopy: Lab

Atomic Spectroscopy

136 Views

article

14.15 : Atomic Fluorescence Spectroscopy

Atomic Spectroscopy

227 Views

See More

JoVE Logo

Privacy

Terms of Use

Policies

Research

Education

ABOUT JoVE

Copyright © 2025 MyJoVE Corporation. All rights reserved