Place brain slice punches from healthy and neurodegenerative mice on sample-holding screens.

Flip the screens and insert them into a tissue plate.

Take a pre-calibrated sensor cartridge with sensor probes. Load the compounds—pyruvate, oligomycin, FCCP, and antimycin A—into the injection ports.

Place the cartridge on the tissue plate. Initiate the run in an extracellular flux analyzer. 

The probes measure the slice’s oxygen consumption rate or (OCR).

With the first injection, pyruvate enters the mitochondria, generating electron carriers.

These carriers donate electrons to Complexes I and II of the electron transport chain or (ETC), initiating an electron transfer reaction that increases oxygen consumption and OCR.

Oligomycin inhibits Complex V, halting ATP synthesis and decreasing OCR.

FCCP, a protonophore, disrupts the proton gradient, increasing oxygen consumption and OCR.

Antimycin A binds to Complex III and stops the ETC, minimizing OCR.

Compare the OCR profiles.

A decreased OCR in the neurodegenerative slices indicates mitochondrial impairment.

article

02:32

Measurement of Oxygen Consumption Rate in Acute Mouse Brain Slices

Ähnliche Videos

7 Views

article

07:11

Messung und Interpretation von Sauerstoff-Verbrauch im ganzen fliegen Kopf Segmente

Ähnliche Videos

15.6K Views

article

11:26

Analyse von Sauerstoff-Verbrauch im primären kultivierten Maus neonatale Herzzellen mit einer extrazellulären Flux-Analyzer

Ähnliche Videos

8.6K Views

article

08:10

Messung von Sauerstoff Verbrauchsmengen in intakten Caenorhabditis elegans

Ähnliche Videos

8.9K Views

JoVE Logo

Datenschutz

Nutzungsbedingungen

Richtlinien

Forschung

Lehre

ÜBER JoVE

Copyright © 2025 MyJoVE Corporation. Alle Rechte vorbehalten